CN.
AutoTFF Pilot10
AutoTFF Pilot10
+
  • AutoTFF Pilot10
  • AutoTFF Pilot10

AutoTFF Pilot10 Crossflow System

Automated tangential flow filtration system for process scale-up and small-scale production, with flexible configuration and a maximum flow rate of 10 L/min.
  • Product Introduction
  • Product Parameters
  • Related Documents

    AutoTFF™Pilot10 is a flexible, automated pilot-scale crossflow system suitable for ultrafiltration, microfiltration, and diafiltration of biological products such as antibodies, recombinant proteins, mRNA, nucleic acids, and viral vectors. The flexible configuration allows for manual operation as well as highly automated process control, with a comprehensive data recording system suitable for process scale up and small-scale production. The tubing connections can accommodate both cassette filters and hollow fibers.

     

    Features: 

    ● High-precision low-pulse diaphragm pump, maximum 10 L/min, suitable for filtration areas of 0.1 m² to 2.5 m².

    ● Automatic TMP control, automatic weighing, and replenishment, precise control of concentration factor and wash volume.

    ● Powerful software include functions of equipment control, recipe setup, data recording, and user management, etc.

    ● Preset common methods with multiple steps that can run sequentially for high automation.

    ● Optional conductivity and pH for more comprehensive process monitoring.

  • General Parameters

    Dimensions(LxDxH,mm)

    945 x 570 x 750 (filter holder not included)

    0.5m²filter holder:420 X 242 X 420

    Power Supply

    240VAC, 50-60Hz

    Weight

    86kg(include 0.1m²filter holder),
    0.1m²filter holder:9kg

    0.5m²filter holder:31kg

    Power

    300W

    System Language

    Chinese, English

    Ambient Temperature

    4-50°C

     

    System Parameters

    Filtration area

    0.1m²- 2.5m²(The specific filtration area can be varied with different filters)

    Maximum feed pressure

    4bar

    Minimum recirculating volume

    300ml

    Feed pump flow rate

    0-10L/min

    Transfer pump flow rate

    0-3000ml/min

    Recirculation vessel volume

    10L

    Pressure sensor

    Detection range: 0-5 Bar

    Permeate flowmeter

    non-contact ultrasonic flow meter, accuracy ±2%, maximum flow rate 3000 ml/min

    Conductivity sensor(Optional)

    Conductivity detection range: 0.001-999.99ms/cm

    Temperature detection range: 0-140℃

    Ph(Optional)

    Detection range: 0-14

    Retentate and filtration weighing

    0-20 kg, accuracy 10 g.

    Filter holder

    Optional 0.1 m² clamp (can hold 1-5 pieces of 0.1 m² cassette filter), made of 316L stainless steel, with an internal surface Ra ≤ 0.4 μm.

    Optional 0.5 m² clamp (can hold 1-5 pieces of 0.5 m² cassette filter), made of 316L stainless steel, with an internal surface Ra ≤ 0.4 μm.

    Optional one hollow fiber support (can hold hollow fibers with a diameter of 1-5 cm).

    Tube&Connector

    Tri Clamp(1/2”)

    Tube

     

    316L stainless steel

    Silicone tubing

    Silicone braided tubing

    PEEK (Polyether ether ketone)

    Control system

     

    13.3" touchscreen

    Industrial all-in-one computer

    Windows 10 operating system

     

     

  • Crossflow System-AutoTFF Pilot10.pdf   

Contact Us

Products of Interest (Multiple Selection)
Services Needed (Multiple Selection)
Submit
%{tishi_zhanwei}%

Inscinstech Co., Ltd.

Provide integrated solutions for biopharmaceutical CMC from upstream and downstream processes to laboratory analysis.

Contact Us